Home· · ·Techcomp college· · ·Application notes
Application notes
Precisa MA08105 – Moisture Determination of Cement
Application Note: Moisture Determination of Cement
This information is intended as a guideline for development of a
moisture analysis method and parameter setting. Materials vary throughout the
world and we will be more than happy to test customers’ individual samples.
Precisa Moisture Analysis XM60/EM120 Halogen
*Can also be conducted using XM50, XM60-HR & XM66
Description: |
Cement |
Utensils: |
Aluminum pan |
Sample |
Cement, fine powder |
Preparation: |
Spread thinly on the pan |
Sample weight (+/-10%): |
2g |
Temperature program: |
Standard |
Temperature: |
105 oC |
Stop-Criteria: |
Autostop 2 Digits [mg] / 60 s |
Moisture content: (average of 6 measurements) |
1.41% |
Standard Deviation (+/-): |
0.02% |
Average analysis time: |
3 min |
Reference: |
Precisa Moisture Analyser XM60/EM120