AN0014 – Optimal Quantification of THC in Blood

INTRODUCTION

The number of regular cannabis users in the world is estimated at 183 million people[1]. It is illegal to drive a vehicle whilst under the influence of cannabis. The active ingredient in cannabis is delta9-Tetrahydrocannabinol (Δ9-THC). Common side effects includes the feeling of euphoria, hallucinations, relaxationand even sedation. Post ingestion, Δ9-THC is metabolised into the inactive ingredient Tetrahydrocannabinolic Acid (THCA/ THC- COOH).
Suspected drug drivers are often subjected to a screening test of cannabinoids in saliva. A positive result must then be confirmed by the presence of Δ9-THC in a blood sample; by gas chromatography. Quantification of THCA can indicate long term use of the drug. Drug detection limits are set by government laws and vary from country to country. The limits in the United Kingdom and Canada is 2ng/mL whereas in France the limit is 1ng/mL.

 

EXPERIMENTAL

A SCION 456 GC coupled with the SCION Single Quad Mass Spectrometer was configured for cannabinoid analysis in blood.

Calibration standards were prepared with a range of 0.5ng/mL to 10ng/mL for Δ9-THC and 2.5ng/mL to 50ng/mL for THC-COOH, with 20μL of internal standard (Δ9-THC-D3 and THC-COOH-D3).

Biological samples were prepared using the following method. 20μL of internal standard was added to 1mL of blood sample. 200μL of acetic acid (10%) and 5mL of hexane/ethyl acetate (9-1 V/V) were added before being centrifuged for 10 minutes at 4000rpm. The organic phase was removed and evaporated at 50°C. The dry extract was then derivatised with 50μL of BSTFA+1%TCMS for 40 minutes at 80°C. 2μL was injected into the split/splitless injector.


Analytical conditions can be found in Table 1.
Table 1. Analytical Conditions of the GC-MS                     


Conditions

S/SL Injector

280°C, splitless, 2μL

Column

SCION-5MS 15m x 0.25mm x 0.25μm

Oven

50°C (hold 0.10min)

15°C/min to 260°C (15.30 mins) 20°C/min to 300°C (18 mins)

Carrier Gas

Helium, 1mL/min

Source

250°C

Transfer Line

310°C

Mode

Positive, 70eV

THC Ions

371, 386

THC-COOH Ions

371, 473

 
RESULTS
The calibration curves for both THC and THC-COOH can be found in Figures 1 and 2. The linearity for both cannabinoids were excellent.

Figure 1. Δ9 THC Calibration Curve



Figure 2
. THC-COOH Calibration Curve

The sensitivity of the SCION SQ MS is excellent, thus exceeding the lowest quantification limit set in France (1ng/mL) , for Δ9-THC. The limit of detection for Δ9-THC is 0.2ng/mL with the limit of quantification 0.5ng/mL whereas the limit of quantitation of THC-COOH is 2.5ng/mL.

Reproducibility (n=10) and repeatability (n=5) of the system were tested using consecutive injections of Δ9-THC at both 0.20ng/mL and 0.5ng/mL plus THC-COOH at 2.5ngmL. Analysing the sample spiked at the limit of quantification shows the high reliability and robustness of the results. Table 2 details the performance of the SCION GC-MS.

Table 2. Reproducibility and repeatability of Δ9-THC and THC-COOH


Analyte

n

Concentration (ng/mL)

RSD (%)

Δ9-THC

10

0.5

6.8

Δ9-THC

5

0.20

3.8

THC-COOH

5

2.5

5.5

 

Figure 3 shows the chromatogram from seven consecutive injections, of 0.5ng/mL Δ9-THC, which were then overlaid in the Mass Spectrometry Work Station software.


Figure 3
. Overlay of 7 consecutive injections of 0.5ng/mL Δ9-THC

A blood sample was obtained from a cannabis smoker. The cannabinoids were extracted from the plasma and ran using the same method. Figure 4 shows the total ion chromatogram (TIC) of both Δ9- THC and THC-COOH.


Figure 4
. TIC of blood sample from cannabis user


CONCLUSION

The analysis of cannabinoids from biological matrices is made easy with the SCION GC-MS. The highly sensitive and robust instrument has exceptional limit of detection for both Δ9-THC and THC-COOH. With an LOD of 0.2ng/mL and LOQ of 0.5ng/mL, for Δ9-THC, these limits exceed the requirements specified by the laws in UK, Canada and France. Additionally, the cannabinoid analyser is capable of detecting THC- COOH at concentrations as low as 2.5ng/mL, with excellent repeatability for both cannabinoids.


REFERENCES

[1]  United  Nations  Office  on  Drugs  and  Crime  (2017). Statistics and Trends in Illicit Drugs. [online] United Nations, p.1. Available at: https://www.unodc.org/wdr2017/field/WDR17_Fact_sheet.pdf [Accessed 11 Dec. 2018].


More information about Scion GCMS please visit http://hk.techcomp.com.hk/EN/product/content/?MenuID=23&ContentID=1524&CID=277

Online PDF reading

Online PDF reading

top
Micro-blog Sina Facebook

Contact UsMore

Hong Kong
  • 2606, 26/F., Tower 1, Ever Gain Plaza, 88 Container Port Road, Kwai Chung, N.T., Hong Kong
  • +852-27519488 / WhatsApp/WeChat HK: +852-8491 7250
  • techcomp@techcomp.com.hk
Singapore
  • 2 International Business Park, #09-06 Tower 1 The Strategy, Singapore 609930
  • +65-62678921
  • techcomp@techcomp.com.sg

powered by QianYe

All Right Reserved 1988-2025©TECHCOMP GROUP
× Sweep The Concern Us