Home· · ·Techcomp college· · ·Application notes
Application notes
Analysis of Diffuser Panel and Reflector of LED Light Bulb
Low electricity consuming LED light bulbs are drawing attention because of its high energy-saving effect. A diffuser panel to reduce the intense light from the light source and a reflector for the efficient use of light are used for a LED light bulb.
In this note, the UH4150 installed with a transmission holder (close contact), which allows the measurement the light including the diffused light, and a diameter 60mm full integrating sphere was used to measure the transmission spectrum of the diffuser panel in a LED light bulb. The reflectance spectrum of the reflector of the LED light bulb was also measured by using the micro total reflectance measurement system (special order), which allows the measurement even for the micro size of about diameter 5mm.
Transmission Spectrum Measurement for LED Light Bulb Diffuser Panel LED
Analysis Condition
Instrument: Model UH4150 Spectrophotometer
Measurement wavelength range: 300~800nm
Scan speed: 300nm/min
Slit: 8nm
Sampling interval: 1nm
Accessory:
· Dia. 60 full-sphere integrating sphere accesory
· Transmittance holder (close contact)
Transmission Spectrum
Analysis Image
Total Reflectance Spectrum Measurement of LED Light Bulb Reflector
Analytical Conditions
Instrument: Model UH4150 Spectrophotometer
Measurement wavelength range: 300~800nm
Scan speed: 300nm/min
Slit: 8nm
Sampling interval: 1nm
Standard reflective material: Aluminum plane mirror
Accessory
· Dia. 60 Standard Integrating Sphere (for total reflectance)
· Micro Total Reflector Measurement System (special order)
Reflectance Spectrum
For more information on Hitachi UV-VIS-NIR, please visit Hitachi High-Tech Science website
https://www.hitachi-hightech.com/global/products/science/ai/
UH4150 Product Information
http://hk.techcomp.com.hk/EN/product/content/?MenuID=22&ContentID=1530&CID=221