What are Absorption, Excitation and Emission Spectra?

What are Absorption, Excitation and Emission Spectra?

 
In this blog post, the differences between molecular absorption, excitation and emission spectra are explained.

Absorption Spectra

Figure 1: Absorption spectrum of anthracene in cyclohexane measured using the FS5 Spectrofluorometer. Experimental parameters: Δλ = 1 nm.


Absorption spectra (also known as UV-Vis spectra, absorbance spectra and electronic spectra) show the change in absorbance of a sample as a function of the wavelength of incident light (Figure 1), and are measured using a spectrophotometer. Absorption spectra are measured by varying the wavelength of the incident light using a monochromator and recording the intensity of transmitted light on a detector. The intensity of light transmitted through the sample, ISample, (such as an analyte dissolved in solvent) and the intensity of light through a blank, IBlank(solvent only) are recorded and the absorbance of the sample calculated using:

The absorbance is linearly proportional to the molar concentration of the sample; which enables the concentration of the sample to be calculated from the absorption spectrum using the Beer-Lambert Law.

Figure 2: Schematic of the measurement of absorption spectra in a spectrophotometer.

Excitation Spectra

Figure 3: Fluorescence excitation spectrum of anthracene in cyclohexane measured using the FS5 Spectrofluorometer. Experimental parameters: λem = 420 nm, Δλem = 1 nm, Δλex = 1 nm.


Fluorescence excitation spectra show the change in fluorescence intensity as a function of the wavelength of the excitation light (Figure 3), and are measured using a spectrofluorometer. The wavelength of emission monochromator is set to a wavelength of known fluorescence emission by the sample, and the wavelength of the excitation monochromator is scanned across the desired excitation range and the intensity of fluorescence recorded on the detector as a function of excitation wavelength. If the sample obeys Kasha’s Rule and Vavilov’s Rule then the excitation spectrum and absorption spectrum will be identical (compare Figures 1 and 3). Excitation spectra can therefore be thought of as fluorescence detected absorption spectra.

Figure 4: Schematic of the measurement of excitation spectra in a spectrofluorometer.

 

Emission Spectra

Figure 5: Fluorescence emission spectrum of anthracene in cyclohexane measured using the FS5 Spectrofluorometer. Experimental parameters: λex = 340 nm, Δλex = 1nm, Δλem= 1 nm


Fluorescence emission spectra show the change in fluorescence intensity as a function of the wavelength of the emission light (Figure 5), and are measured using a spectrofluorometer. The wavelength of excitation monochromator is set to a wavelength of known absorption by the sample, and the wavelength of the emission monochromator is scanned across the desired emission range and the intensity of the fluorescence recorded on the detector as a function of emission wavelength.


Figure 6: Schematic of the measurement of emission spectra in a spectrofluorometer.


FLS1000


FS5

 

DS5






Online PDF reading

Online PDF reading

top
Micro-blog Sina Facebook

Contact UsMore

Hong Kong
  • 2606, 26/F., Tower 1, Ever Gain Plaza, 88 Container Port Road, Kwai Chung, N.T., Hong Kong
  • +852-27519488 / WhatsApp/WeChat HK: +852-8491 7250
  • techcomp@techcomp.com.hk
Singapore
  • 2 International Business Park, #09-06 Tower 1 The Strategy, Singapore 609930
  • +65-62678921
  • techcomp@techcomp.com.sg

powered by QianYe

All Right Reserved 1988-2025©TECHCOMP GROUP
× Sweep The Concern Us